LED Measurement System
User Manual
22
Ossila Limited © 2022
I.
Resistance can be calculated using: R = V / I
II.
For the 200 mA current range the calculated resistance should be between 98 and
102 kΩ (within 2% of the resistor value).
III.
For all other ranges the calculated resistance should be between 99 and 101 kΩ
(within 1% of the resistor value).
Figure 10.3.
Example measurement of resistor test device using the 200 μA current range.
Содержание T2005
Страница 27: ......