Orthofix Centurion POCT Скачать руководство пользователя страница 3

3

3.  Use a soft cloth or a soft plastic bristle brush to remove any visible soil from the 

instruments prior to manual or automated cleaning. Use a soft plastic bristle brush 

or a pipe cleaner to remove soil from any inner lumens. You can also use a syringe (if 

appropriate) for hard to reach areas.

4.  Enzymatic detergent should be used for manual and automated cleaning. All enzymatic 

detergents should be prepared at the use dilution and temperature recommended by 

the manufacturer. Softened tap water may be used to prepare the enzymatic detergents. 

Use of recommended temperatures is important for optimal performance of enzymatic 

detergent.

Manual Cleaning:

1.  Completely submerge instruments in an enzymatic detergent and allow to soak for 20 

minutes. Use a soft-bristled, nylon brush to gently scrub the device until all visible soil has 

been removed. Particular attention must be given to crevices, lumens, mated surfaces, 

connectors and other hard-to-clean areas. Lumens should be cleaned with a long, narrow, 

soft-bristled brush (i.e. pipe cleaner brush).

2.  Remove the instruments from the enzymatic detergent and rinse in tap water for a 

minimum of 3 minutes. Thoroughly and aggressively flush lumens, holes and other 

difficult to reach areas.

3.  Place prepared cleaning solution in a sonication unit. Completely submerge device in 

cleaning solution and sonicate for 10 minutes.

4.  Rinse instrument in purified water for at least 3 minutes or until there is no sign of blood 

or soil on the device or in the rinse stream. Thoroughly and aggressively flush lumens, 

holes and other difficult to reach areas.

5.  Repeat the sonication and rinse steps above.

6.  Remove excess moisture from the instrument with a clean, absorbent and non-shedding 

wipe.

7.  Inspect the instruments for visible soil.

8.  If visible soil is noted, repeat the steps listed above.

Automated Cleaning:

1.  Completely submerge the instruments in an enzymatic detergent and allow to soak and 

sonicate for 10 minutes each. Use a soft nylon bristled brush to gently scrub the device 

until all visible soil has been removed. Particular attention must be given to crevices, 

lumens, mated surfaces, connectors and other hard to clean areas. Lumens should be 

cleaned with a long, narrow, soft nylon bristled brush (i.e. pipe cleaner). Use of a syringe 

or water jet will improve flushing of difficult to reach areas and closely mated surfaces.

2.  Remove instruments from the cleaning solution and rinse in purified water for a minimum 

of 1 minute. Thoroughly and aggressively flush lumens, blind holes and other difficult to 

reach areas.

3.  Place instruments in a suitable washer/disinfector basket and process through a standard 

instrument washer/disinfector cleaning cycle. 

4.  Orient instruments into the automated washer’s carriers as recommended by the washer 

manufacturer.

5.  The following minimum parameters are essential for thorough cleaning. 

a.  2 minute prewash with cold tap water 

b.  1 minute prewash with hot tap water

c.  2 minutes detergent wash with hot tap water (64-66°C/146-150°F)

d.  1 minute hot tap water rinse

e.  2 minute thermal rinse with purified water (80-93°C/176-200°F)

f.  1 minute purified water rinse (64-66°C/146-150°F)

g.  7 to 30 minute hot air dry (116°C/240°F) 

6.  Inspect the instruments for visible soil.

7.  If visible soil is noted, repeat the above listed steps until no visible soil is noted.

Note: 

Certain cleaning solutions such as those containing caustic soda, formalin, 

glutaraldehyde, bleach, and/or other alkaline cleaners may damage instruments. These solutions 

should not be used.

Note: 

Visually inspect instruments after cleaning and prior to each use. Discard or return to 

Orthofix any instruments that are broken, discolored, corroded, have cracked components, pits, 

gouges, or are otherwise found defective. Do not use defective instruments. 

Instrument End of Life Determination:

Do not reuse Single Use instruments. Visually inspect the reusable instruments to determine if 

the instrument has reached end of life. Orthofix reusable instruments have reached End of Life 

when:

1.  Instruments show signs of damage such as binding, bending, breakage, overt signs of 

wear and/or any other conditions which may impact the devices safe and effective use.

2.  Instruments intended for cutting bone and/or tissue (e.g. tap, rasp, curette, rongeur) – 

when any of the cutting surfaces show signs of wear such as nicks, abrasions or 

otherwise dulled cutting surfaces.

3.  Instruments that interface with other devices (e.g. implants, instruments, handles) – when 

the mating feature binds, fails to attach or fails to hold the device securely. The instrument 

function should be verified prior to each use.

4.  Do not use instruments which reached End of Life. Discard End of Life instruments per 

your hospital procedure or return to Orthofix for disposal. 

Sterilization:

Sterilization in Orthofix Cases with Blue Wrap:

The Centurion POCT System implants and instruments are supplied NON-STERILE. Prior to use, 

all implants and instruments should be place in the appropriate Orthofix case which will be 

wrapped in a FDA cleared sterilization wrap and placed in the autoclave for sterilization by the 

hospital using one of the following recommended cycles:

Method: Steam

or:

Method: Steam

Cycle: Gravity

Cycle: Pre-vacuum

Temperature: 270°F (132°C )

Preconditioning: Per manufacturer settings

Exposure time: 15 minutes 

Temperature: 270°F (132°C)

Drying time: 30 minutes

Exposure time: 4 minutes

Drying time: 30 minutes

Sterilization in Rigid Sterilization Containers:

When using rigid sterilization containers, clean, inspect and prepare the rigid sterilization

container according to the manufacturer’s instructions.

Select the appropriate rigid sterilization container with either a filtered or solid bottom to 

properly enclose the Orthofix case(s) or caddies (recommended 23¼” long x 11¼” wide 

container). The following sterilization cycle has been validated:

Method: Steam

Cycle: Pre-vacuum

Temperature: 270°F (132°C)

Preconditioning: Per manufacturer settings

Exposure time: 4 minutes

Drying time: 30 minutes

Validation and routine monitoring should be performed as per ANSI/AAMI ST79 Comprehensive 

guide to steam sterilization and sterility assurance in health care facilities. Other cycles may be 

used as long as they comply with the above practices and provide a sterility assurance level of 

10

-6

.

Physician Information:

Patient Selection:

Patient selection is an extremely important factor in the success of implant procedures. It is 

important that the candidates be carefully screened and the optimal therapy selected.

Preoperative:

1.  Only patients that meet the criteria described in the indications should be selected.

2.  Patient conditions and/or predispositions such as those addressed in the aforementioned 

contraindications should be avoided.

3.  Correct selection of the implant is extremely important.

4.  Use care in handling and storage of implant components. Cutting, bending, or scratching 

the surface of metal components can significantly reduce the strength and fatigue 

resistance of the implant system and should be avoided. These, in turn, may cause cracks 

and/or internal stresses that are not obvious to the eye and may lead to fracture of the 

components. Inspection should be made to determine if components have been damaged 

during storage or previous procedures.

5.  An adequate inventory of implant sizes should be available at the time of surgery.

6.  Certain special surgical instruments are required to perform this surgery. Review of the 

use and handling of these instruments is very important.

7.  PRE-OP PLANNING – Use of cross sectional imaging (i.e., CT and/or MRI) for posterior 

cervical screw placement is recommended due to the unique risks in the cervical spine. 

The use of planar radiographs alone may not provide the necessary imaging to mitigate 

the risk of improper screw placement. In addition, use of intraoperative imaging should 

be considered to guide and/or verify device placement, as necessary.

Intraoperative:

1.  Whenever possible, use pre-cut rods of the length needed. The rods should not be 

repeatedly or excessively bent any more than absolutely necessary. The rods should not 

be reverse bent in the same location. Use great care to insure that the implant surfaces 

are not scratched or notched in any way. If the rods are cut to length, they should be cut 

in such a way as to create a flat, non-sharp surface perpendicular to the midline of the 

rod. 

2.  The use of two rods and cross connecting the rods will provide a more rigid construct.

3.  The placement of screws should be checked radiographically prior to assembly of the rod 

construct.

4.  Care should be taken when positioning the implants to avoid neurological damage.

5.  To facilitate proper fusion below and around the location of the instrumentation, a bone 

graft should be used.

6.  Confirm that the rods are fully seated in the bottom of the screw head. Rods that are not 

fully seated may prevent the device from locking together. 

7.  Before closing the soft tissues, all of the set screws should be tightened firmly with a 

torque wrench and set screw driver according to the operative technique. Recheck the 

tightness of all screws to make sure that none loosened during the tightening of the other 

set screws. Failure to do so may cause loosening of the other components.

8.  Bone cement should not be used since this material will make removal of the component 

difficult or impossible. The heat generated from the curing process may also cause 

neurological damage and bone necrosis.

Postoperative:

1.  Detailed instructions on the use and limitations of the implant should be given to the 

patient. The patient must be made aware of the limitations of the implant. Physical 

activity and load bearing have been implicated in premature loosening, bending, or 

fracture of internal fixation devices.

2.  Periodic X-rays for at least the first year postoperatively are recommended for close 

comparison with postoperative conditions to detect any evidence of changes in position, 

nonunion, loosening, and bending or cracking of components.

3.  Surgical implants must never be reused. Any retrieved devices should never be reused in 

another surgical procedure. The retrieved parts should be handled and disposed of in such 

a manner as to ensure that reuse is not possible.

4.  To allow the maximum potential for a successful surgical result, the patient or device 

should not be exposed to mechanical vibration that may loosen the device construct.

5.  These implants are temporary internal fixation devices. Internal fixation devices are 

designed to assist in the stabilization of the operative site during the normal healing 

process. After healing occurs, these devices serve no functional purpose and should be 

removed. In most cases removal is indicated because the implants are not intended to 

transfer or support forces developed during normal activities. If the device is not removed 

following completion of its intended use, complication may occur as follows: 

a.  Corrosion, with localized tissue reaction or pain. 

b.  Migration of implant position resulting in injury. 

c.  Risk of injury from postoperative trauma. 

d.  Bending, loosening and/or breakage, which could make removal impractical or 

difficult. 

e.  Pain, discomfort or abnormal sensations due to the presence of the device. 

f.  Possible increased risk of infection. 

g.  Bone loss caused by stress shielding.

Adequate postoperative management to avoid fracture, re-fracture or other complications should 

follow implant removal.

Содержание Centurion POCT

Страница 1: ...PL US Orthofix Inc 05 2021 Orthofix Inc 3451 Plano Parkway Lewisville Texas 75056 9453 U S A 1 214 937 3199 1 888 298 5700 www orthofix com Medical Device Safety Service MDSS Schiffgraben 41 30175 Ha...

Страница 2: ...l defects and internal stress patterns that may lead to fatigue failure 7 The implantation of POCT systems should be performed only by experienced spinal surgeons with specific training in the use of...

Страница 3: ...ons Select the appropriate rigid sterilization container with either a filtered or solid bottom to properly enclose the Orthofix case s or caddies recommended 23 long x 11 wide container The following...

Страница 4: ...e the system s components The system s instruments are organized into trays within the modular case for easy retrieval during surgery These trays also provide protection to the system components durin...

Страница 5: ...ganchos las varillas los adaptadores offset los conectores cruzados las placas los conectores de varilla y los instrumentos se venden sin esterilizar y por lo tanto deben esterilizarse antes de usarse...

Страница 6: ...e 7 Si se observa suciedad visible repita los pasos anteriores hasta que no quede suciedad visible Nota Ciertas soluciones de limpieza como las que contienen sosa c ustica formol glutaraldeh do lej a...

Страница 7: ...e Dolor molestias o sensaciones an malas debido a la presencia del dispositivo f Posible aumento del riesgo de infecci n g P rdida sea causada por el efecto conocido como stress shielding u osteopenia...

Страница 8: ...ntos s o vendidos n o est reis e por isso t m de ser esterilizados antes da utiliza o 4 A incapacidade para se obter a artrodese resultar no eventual afrouxamento e falha da estrutura do dispositivo 5...

Страница 9: ...jidade vis vel Nota Algumas solu es de limpeza como as que cont m soda c ustica formol glutaralde do lix via e ou outros agentes alcalinos podem danificar os instrumentos N o deve utilizar estas solu...

Страница 10: ...normais decorrentes da presen a do dispositivo f Poss vel aumento do risco de infe o g Perda de osso provocada por remo o das tens es normais exercidas sobre o mesmo Ap s a remo o do implante deve ter...

Страница 11: ...n 3 Unsteril Die Schrauben Feststellschrauben Haken St be Distanzst ckadapter Querverbinder Platten Stabverbinder und Instrumente sind im Lieferzustand unsteril und m ssen daher vor Gebrauch sterilisi...

Страница 12: ...s Waschger ts ausrichten 5 F r eine gr ndliche Reinigung sind unbedingt die folgenden Mindestparameter einzuhalten a 2 Minuten Vorw sche mit kaltem Leitungswasser b 1 Minute Vorw sche mit hei em Leitu...

Страница 13: ...ungen sind darauf ausgelegt die Stabilisierung des Operationssitus w hrend des normalen Heilungsvorgangs zu unterst tzen Mit Abschluss der Heilung werden diese Implantate funktionslos und sind zu entf...

Страница 14: ...ords lat raux raccords transversaux plaques raccords de tige et instruments sont vendus non st riles et doivent tre st rilis s avant l utilisation 4 L chec de l arthrod se entra ne le descellement ou...

Страница 15: ...yants alcalins peuvent endommager les instruments Ces solutions ne doivent pas tre utilis es Remarque Inspecter visuellement les instruments apr s le nettoyage et avant chaque utilisation Mettre au re...

Страница 16: ...ons anormales dues la pr sence du dispositif f Augmentation possible du risque d infection g D perdition osseuse caus e par un effet de bouclier anti contrainte Un traitement postop ratoire appropri p...

Страница 17: ...condizioni di non sterilit e devono pertanto essere sterilizzati prima dell uso 4 Il mancato ottenimento di artrodesi comporta nel tempo l allentamento e il cedimento del costrutto del dispositivo 5...

Страница 18: ...ottura segni evidenti di usura e o altre condizioni che potrebbero influire sulla sicurezza e sull uso efficace dei dispositivi 2 Strumenti previsti per il taglio di osso e o di tessuto es maschiatori...

Страница 19: ...rgia La limitazione dimensionale massima aumenta il rischio di complicazioni meccaniche come l allentamento la piegatura o la rottura dei dispositivi Tutte queste complicazioni possono rendere necessa...

Страница 20: ...evit jako nepo kozen mohou v n m b t mal defekty a vzory vnit n ho pnut kter mohou v st k navov zlomenin prost edku 7 Implantaci syst mu POCT by m li prov d t pouze zku en spin ln chirurgov konkr tn v...

Страница 21: ...n pro ez n kost pop tk n nap z vitn k ra ple kyreta t pac kle t kdy jak koli ezac povrch vykazuje zn mky opot eben nap klad vrypy od en nebo jinak ztupen ezn plochy 3 N stroje p il haj c k jin m prost...

Страница 22: ...hojit je flexibiln a ob as se m e zlomit pop degraduje Anatomie lidsk ho t la stanovuje ur it omezen co do velikosti jak hokoli um l ho fixa n ho prost edku pou van ho v chirurgii Omezen maxim ln veli...

Страница 23: ...u it m sterilizova 4 Ak sa nedosiahne artrod za m e to ma za n sledok uvo nenie a rozpad trukt ry pom cky 5 Aplik cia pr li vysok ho kr tiaceho momentu na skrutky m e sp sobi po kodenie z vitov v kost...

Страница 24: ...osti Orthofix Po koden n stroje nepou vajte Ako ur i ukon enie ivotnosti n stroja N stroje ur en na jednorazov pou itie nepou vajte opakovane Zrakom skontrolujte dosiahnutie ivotnosti n strojov ur en...

Страница 25: ...u it pri oper cii chrbtice s kovov implant ty ktor sa pripev uj ku kosti a pom haj pri hojen kostn ch tepov Je dok zan e tieto implant ty prin aj chirurgom cenn pomoc pri lie be kostn ch f zi Tieto po...

Отзывы: