30
5. Secure the green, white and black wires to the underside of the Power Inlet using the 11.5” tie-wrap. Be sure to
also secure the green/yellow wire to the side of the power inlet. Wrap the tie-wrap around the Power Inlet and
all wires, securing the wires between the tie-wrap and the Power Inlet. Tighten the tie-wrap so that the head of
the tie-wrap is located at the bottom left corner of the Power Inlet. Once tight and secure cut the excess tie-
wrap with diagonal wire cutters.
6. Attach the other end of the Retractile Cable to the centrifuge drawer using the same two screws and washers
previously removed.
7. Insert the exposed end of the white wire into the terminal block opening closest to the Main PC Board, and
secure using the flat head screwdriver to tighten the terminal block screw.
8. Insert the exposed end of the black wire into the remaining terminal block opening, closest to the latch
solenoid, and secure using the flat head screwdriver to tighten the terminal block screw.
9. Attach the green wire to the centrifuge drawer ground location by inserting the screw (previously removed)
through the washers first and then through the connector ring at the end of the wire. Secure the screw to the
centrifuge drawer.
10. With the 4" tie-wrap, gather the black, white, and green wires and attach to the orange wire in a manner that
puts no stress on any of the wire connections while at the same time keeps the wires down and away from the
Retractile Cable.
11.5" tie wrap
Black, White, & Green Wires
Green/Yellow Wire
tie wrap head
Содержание MTS Centrifuge 5150-60
Страница 1: ...MTS Centrifuge Model 5150 60 Service Manual PK101 doc OCD CTS part number 50000104 Cat Number 6902119 ...
Страница 2: ......
Страница 6: ...This page intentionally blank ...
Страница 12: ...6 Refer to Section 1 5 Parts List for Item Key Number ...
Страница 17: ...11 2 Equipment Installation Choice of location Unpacking ...
Страница 21: ...15 3 2 Main PCB Schematic ...
Страница 22: ...16 3 3 Display PCB Schematic ...