8.2 Power Inlet / Power Switch Removal
1. Remove all wires and connectors.
2. Remove the two screws.
3. Remove the inlet.
To install a new power inlet:
1. Insert the power inlet into the back panel. (Note: The power switch section of the inlet should face
towards the shorter end of the panel.)
2. Secure the inlet to the back panel using the two screws and blue Loctite.
3. Attach all wires and connectors (see wiring diagram 3.7).
8.3 Retractile Cable Sub-Assembly Removal (Back Panel Side)
1. See Retractile Cable Sub-Assembly Removal section 7.9
To install a new retractile cable:
1. See Retractile Cable Sub-Assembly Removal section 7.9