7.5 Transformer Removal
1. Remove the connector from the main board.
2. Remove the two black wires from the terminal block connector.
3. Loosen and remove the two 8-32 X 5/16 screws and washers.
4. Remove transformer.
To install a new transformer:
1. Secure the transformer to the chassis with the two screws, washers and blue Loctite (Note: The black
wires of the transformer should face the rear of the chassis).
2. Place the connector on the main board.
3. Connect the two black wires to the two terminals of the terminal block connector (Note: It is not
important which black wire is connected to which connector.).
7.6 Latch Mechanism Removal
1. Remove all wires and connectors.
2. Remove the two M4 X 10 PH Phillips screws and washers.
3. Remove latch mechanism.
To install a new latch mechanism:
1. Secure the latch to the chassis using the two screws, washers and blue Loctite.
2. Replace all wires and connectors (see wiring diagram 3.6).
3. Adjust the latch mounting so that the drawer closes and latches with no interference.