29
7.9 Retractile Cable Sub-Assembly Removal
1. Using a flat head screwdriver, loosen the screws on the terminal block that secure the white and black Retractile
Cable wires. Remove the wires from the terminal block.
2. Using a phillips screwdriver, unscrew the ground location screw that connects the green Retractile Cable wire
to the centrifuge drawer. Place the screw and washer aside for re-assembly.
3. Cut the single tie-wrap that attaches the white, black, green, and orange wires together.
4. Using a phillips head screwdriver, remove the two screws and washers that attach the cable hanger to the
centrifuge drawer. Place the screws and washers aside for re-assembly.
5. The back panel plate is now detached from the centrifuge drawer. Take a moment to examine the diagram of
the Retractile Cable's wires to the power inlet, (Figure 7.1 or 7.2).
6. Disconnect the white, black, and green wires from the power inlet (Figure 7.1 or 7.2).
7. Using a phillips head screwdriver, remove the ground location screw that attaches the green/yellow wire to the
back panel plate (Figure 7.1 or 7.2).
8. Using a phillips head screwdriver, remove the single screw and washers that attach the protector plate to the
back panel (Figure 7.1 or 7.2). Place the screw and washers aside for re-assembly. The Retractile Cable with
cable hanger and protector plate should now be completely removed from the centrifuge.
Replacement Instructions - Installing the Retractile Cable:
1. There are two different ends to the retractile cable; one end is attached to the protector plate and the other end is
attached to the cable hanger.
2. Attach the end with the protector plate to the back panel using the phillips head screwdriver to attach the same
screw and washers that were previously removed.
3. Attach the push-on connectors
from the green, black, and white
wires to the power inlet.
4. Attach the green/yellow wire to
the back panel by inserting the
screw (previously removed)
through the washers first and
then though the connector ring at
the end of the wire. Secure the
screw to the back panel.
Black Wire
White Wire
Figure 7.1:
All Serial #’s XX-XX-0001 through XX-XX-2230
Figure 7.2:
All Serial #’s from XX-XX-2230 and after
Green Wire
Green/Yellow Wire
Ground Location
Black Wire
White Wire
Green Wire
Green/Yellow Wire
Ground Location
Содержание MTS Centrifuge 5150-60
Страница 1: ...MTS Centrifuge Model 5150 60 Service Manual PK101 doc OCD CTS part number 50000104 Cat Number 6902119 ...
Страница 2: ......
Страница 6: ...This page intentionally blank ...
Страница 12: ...6 Refer to Section 1 5 Parts List for Item Key Number ...
Страница 17: ...11 2 Equipment Installation Choice of location Unpacking ...
Страница 21: ...15 3 2 Main PCB Schematic ...
Страница 22: ...16 3 3 Display PCB Schematic ...