8
RPM Display
RPM does not display while
Check opto sensor adjustment.
centrifuge is spinning.
Remove front panel cover and check ribbon cable
connection.
Check connections of opto assembly on tach PC board.
Check connection of tach PC board to main PC board.
Remove electronics cover and check ribbon cable
connection.
Replace
opto
assembly.
Replace
main
PC
board.
RPM flashes erroneous numbers
Check opto sensor adjustment.
Check connections of opto assembly to tach PC board.
Replace
opto
assembly.
Replace
main
PC
board.