9. Removal Instructions - Cabinet
9.1 Tachometer Cap Removal
The tachometer cap is designed to be removed and replaced easily.
To remove it grasp it on two sides and pull up. If this is difficult, reach inside the cabinet and pop the cap out from
the inside.
9.2 Rubber Feet Removal
The rubber feet are attached to 8-32 screw studs which are screwed into the bottom of the cabinet.
9.3 Rubber Seal Removal
The rubber seal is attached to the cabinet using eight 4-40 nuts around the inside edges. (Note: The bubble part of
the seal should face towards the front of the instrument.)