[REV. 20210125] OW_ORBmax_BA_882060201_01__INH-02_EN
Orbitalum Tools GmbH
D-78224 Singen www.orbitalum.com
35
ORBMAX Operation
eNGLISH
After the ”
Start logging
” key (
9
) has been pressed, a welding seam number (max. 6 digits) under which a measurement is to be
stored afterwards on the SD card has to be entered first.
8.2.2 ”Start pump” key
The ”
Start Pump
”
(
10
) key is used to switch the pump on and
off and the actual measuring process is started.
The
wheel
(
11
) on the main screen rotates while the pump
is running. When the pump is standing still, the wheel also
stands still.
The
pump output
(
12
) is only displayed during a measure-
ment and can be preset in the menu under ”
Pump Power
”
(see chap. 8.2.3).
Presettings for a measurement/measuring series are carried
out in the ”
Menu
” (
13
) (see chap. 8.2.3).
13
11
10
12
8.2.3 ”Menu” key
The ”
Menu
” (
13
) – see figure above – is used to access the
main menu.
Several setting options are available:
The menu language can be set to German, English, Italian,
French, Spanish or Polish in the ”
Language
” (
14
) menu item.
In the case of forming applications with a low volume flow
the pump power can be reduced (100% = approx. 3 l/min)
by means of the ”
Pump Power
” (
15
) key (selection options:
100%/80%/60%/40%/20%).
14
15
16
17
18
19
20
21
22
The software version number installed on the device is displayed under ”
Version
” (
16
).
The alarm can be activated or deactivated by means of the ”
Alarm
” (
17
) menu. As long as the measured value (in ppm) lies over
the set alarm value, a red bar is displayed on the main screen and, if this is set, a warning sound is issued (”
Alarm On with Beeper
”).
If the value drops below the alarm value, the bar turns green and the beeper stops.
The value for the ”
Log interval
” (
18
) and the ”
Alarm value
” (
19
) of a measuring series can be set via the respective menu items.
The ”
Back
” key (20) is used to return to the main display.
”
Date
” (
21
) and ”
Time
” (
22
) can be set or changed by means of the respective keys.
NOTe!
The current date, the time (Central European Time) as well as the menu language (German)
are preset in the factory before the ORBmax is delivered.