Servicing, maintenance, troubleshooting
eNGLISH
9. SeRVIcING, MAINTeNANce, TROUBLeSHOOTING
NOTe!
Some of the tasks named are heavily dependent on use and ambient conditions. The
stated cycles are minimum values. In individual cases, different maintenance cycles are
possible. To ensure the device safety, have the calibration performed annually by autho-
rized service points with VDE testing. Should the device not function as described above, it
must also be sent to an authorized service point.
DANGeR!
Danger of death by electric shock!
Non-observance could result in death or critical injury.
X
Switch the device off and unplug the power plug before carrying out transportation,
cleaning, maintenance and repairs.
DANGeR!
Risk of electric shock due to poor electrics!
Fatal electric shock.
X
Switch the device off and unplug the power plug before carrying out transportation,
cleaning, maintenance and repairs.
X
Repair and maintenance work on the electrical equipment may only be carried out by
a qualified electrician.
9.1
Maintenance
INTeRVAL
AcTIVITY
Depending on the degree of soiling Replacing the filter insert:
X
Open the filter at the measuring hose.
X
Replace the filter insert.
X
Close the filter again.
Annually
X
The oxygen sensor is maintenance-free. However, annual calibration is recommended,
as for all devices. Please send the device directly to the branch responsible for you.
9.2
What to do if? – General troubleshooting
PROBLeM
POSSIBLe cAUSe
ReMeDY
Device cannot be switched on.
Power plug not plugged in correctly.
X
Check whether the device or the power
adapter is plugged in correctly.
Plug power supply unit defective.
X
Replace the plug power supply unit.
Oxygen value does not drop during
measuring, but remains at the
dummy value of ”999”.
Hose connections not plugged correctly.
X
Check hose connections. If necessary,
replug.
Measuring hose is damaged and leaks.
X
Check the measuring hose for damage. If
necessary replace the hose.