The Laser, ADC, Output, Modulation, and Maintenance menu
The right side of the screen is reserved for the
Laser
,
ADC
,
Output
,
Modulation
, and
Maintenance
menu.
To scroll through the menus, move the cursor to the section right of the graph area, and press the left or
right arrow (see
).
Figure 51:
Laser (a), ADC (b), Output (c), Modulation (d) and Maintenance menu (e) of the OP1550 interferometer.
The ‘
Laser
’ menu displays the laser controls. Always operate channel 86 of the laser: this corresponds to
1530 nm. The minimum laser power is 6.31mW and the maximum one is 25mW.
Caution:
Please consider increasing laser power to a maximum one when measuring in
the medium as it will increase your signal-to-noise ratio (see the video:
). However, for the wavelength scan in the air, please
decrease the laser power as there might be too much signal which would result in a
saturated signal (flat line after wavelength scan).
Figure 50:
Datapoint changing its position after the tip is getting in contact with a surface: probe out of contact (a), noise caused
by probe moving during automated find surface procedure or by manually moving the Z-stage (b), probe in contact with surface
resulting in datapoint position change (c).