38
optics11life.com
differ depending on PID settings
set in the ‘Maintenance’
. Therefore, initial tests need to be performed
to evaluate system performance for the specific sample/probe combination. Get in contact with
[email protected] to get more information and training. The number of periods should be chosen
so that oscillations take at least 1s, for instance when using 0.5 to 5Hz, the number of periods should be
5 while for higher frequencies, the number of periods should match the frequency, e.g., 75Hz - 75 periods.
Table 4: Parameters range for DMA profile.
Min. initial relaxation time
5s
Min. relaxation time
2s
Min. amplitude
2x noise level
Lowest frequency possible
0.01 Hz
Highest frequency possible
20 Hz closed loop, 75Hz open loop
Min. oscillation time
1s
Figure 30:
Configuration of DMA profile.
Dynamic Mechanical Analysis (DMA) uses a cyclic motion with frequency,
𝑓
, while controlling indentation-
depth or load with an amplitude of
ℎ
0
or
𝐹
0
, respectively, if used in closed-loop. In an open-loop, only the
displacement of piezo is controlled, thus,
ℎ
0
and
𝐹
0
will depend on the relationship between sample and
probe stiffnesses. The frequency-dependent storage modulus,
𝐸′
, and loss modulus,
𝐸′′
, which represent
elastic and viscous components respectively, are calculated with the following equations:
𝐸
′
(𝑓) =
1
2
𝐹
0
ℎ
0
cos(𝛿)
(1 − 𝜈
2
)
√ℎ𝑅