6212C Manual Version 15_22 Rev 07-19.docx
Page 50 of 53
@23,xxxxx#<cr-lf>, Set/Query Head Serial Number
Note: Factory Set
Set Command:
@23,xxxxx#<cr-lf>, Set/Query Head Serial Number
Query Command:
@23#<cr-lf>
Query Response:
@23,C11xxxxx,#<cr-lf>
@26,b,xxx.xxV#<cr-lf>, Set/Query resistor values
Note: Factory Set
Set command:
@26,b,xxx.xxV#<cr-lf>
b is the Bank Number. Range is 1-5.
xxx.xx is the resistor value. Default is 1.0K.
V is the multiplier, either K for kilohm or M for Megohm.
Values are in ASCII with a required 2 digits to the right of the decimal
point and 1 to 3 digits to the left.
Example
Set bank2 to 3.0K resistor from default 1.0K
@26,2,3.00K#<cr-lf>
Query
@26#<cr-lf>
Response
@26#<cr-lf>
001.00K#<cr-lf>
003.00K#<cr-lf>
001.00K#<cr-lf>
001.00K#<cr-lf>
001.00K#<cr-lf>
#<cr-lf>