6212C Manual Version 15_22 Rev 07-19.docx
Page 45 of 53
max_detector is the integer value of the detector with the maximum voltage on it. This value will always be 2
digits and can range from 01 to 35. The location of the maximum detector voltage is useful for determining
vertical alignment on smooth surfaces only. Refer to the alignment section in the manual.
sum_voltages is the sum of all voltages from detectors 01 to 35. The value will be 7 characters in length
(including the decimal point character) with padding zeros on either end when necessary, and always followed
by a comma.
@04#<cr-lf> Specular Values Request
Response
@04<cr-lf>
00.1234,00.1234<cr-lf>
Spec sums Rough, Smooth used for Ra calculation
06,00.1234<cr-lf>
Location, Sum of 3 max used to find specular
07,0.1234<cr-lf>
Max Detector Loc, Max Det Value within "Sum of 3 max"
#<cr-lf>
@05 Request Ra Value
Returns Ra values when the start input pin is activated, waiting for stop input.
Stop trigger input is Digital IN 2 or Pin 2 of connector J7. Stop conversions when pin is activated.
@05#<cr-lf> Returns a single @05 response. If continuous mode was running, it now quits
Response
@05,000.1234,001.1234,ok,dd,12.3456,#<cr-lf>
Ra rough, Ra smooth, Error Code, Max Detector, Sum Voltage of Detectors 1-37
@05,00#<cr-lf> Returns continuous stream of Ra values when the start input pin is activated until stop input
pin is activated.
Response
@05,000.1234,001.1234,ok,dd,12.3456,#<cr-lf>
@05,000.1234,001.1234,ok,dd,12.3456,#<cr-lf>
@05,000.1234,001.1234,ok,dd,12.3456,#<cr-lf>
When the stop input is activated, the Ra stream will stop.
Ra rough, Ra smooth, Error Code, Max Detector, Sum Voltage of Detectors 1-37