Message Types:
@02 Request Ra Value
@02#<cr-lf> Returns a single @02 response. If continuous mode was running, it now quits
Response
@02,00.1234,01.1234,ok,dd,12.3456,#<cr-lf>
Ra rough, Ra smooth, Error Code, Max Detector, Sum Voltage of Detectors 1-37
@02,00#<cr-lf> Returns continuous @02 response. Will continue making measurements until a
@02#<cr-lf> command stops it.
Response
@02,00.1234,01.1234,ok,dd,12.3456,#<cr-lf>
@02,00.1234,01.1234,ok,dd,12.3456,#<cr-lf>
@02,00.1234,01.1234,ok,dd,12.3456,#<cr-lf>
continuous until @01#<cf-lf> received
Ra rough, Ra smooth, Error Code, Max Detector, Sum Voltage of Detectors 1-37
@02,dd#<cr-lf> Returns continuous @02 response. Will continue making measurements until "dd"
number of measurements have occurred (dd = 01 to 99).
Response
@02,00.1234,01.1234,ok,dd,12.3456,#<cr-lf>
@02,00.1234,01.1234,ok,dd,12.3456,#<cr-lf>
repeats "dd" times
@02,00.1234,01.1234,ok,dd,12.3456,#<cr-lf>
Response Details:
The response from the interconnect board will be the rough then the smooth roughness values. The 2 roughness
values will always be 7 characters in length (including the decimal point character) with padding zeros on either
end when necessary, and always followed by a comma.
Ra rough, Ra smooth, Error Code, Max Detector, Sum Voltage of Detectors 1-37
The error codes will be 2 letters followed by a comma.
ok
–
for measurement completed successfully.
tc
–
for smooth measurement “too close”
tf
–
for smooth measurement “too far”