SPECIFICATIONS
Dimensions
Outside
30” wide x 46.5” deep x 72” high (762 x 1180 x 1828 mm)
Weight
375 lbs. (170 kg)
Functional
Operating Power / Current 110V: 785W / 7.2 A, 240V: 755W / 3.3 A
Standby Power / Current
110V: 19W / 175 mA, 240V: 21W / 158 mA
Input Voltage
110-240VAC, 50/60 Hz (voltage fluctuation ±10%)
Transient over voltages allowed up to category II
Temporary over voltages allowed to occur on mains supply
Power Connection
Connect to branch circuit rated 20A max
Fuse Protection
5x20mm, Slow Blow, 10A, 250V
Note:
Disconnect main power supply before replacing fuse.
Débranchez l’alimentation principale avant de remplacer le
fusible.
Environmental
Operating
41°F (5 °C) to 104 °F (40 °C) temperature
Maximum relative humidity 80% for temperatures up to 88°F
(31°C), decreasing linearly to 50% relative humidity at 104° F (40°C)
Note: Under 41°F (5°C) temperature must run warm-up cycle before
disinfecting. See Operating section.
Storage
-20 °F (-28 °C) to 140 °F (60 °C) temperature, 10–95% humidity
Note: Allow unit to reach operating temperature before use.
Pollution Degree 2
UV-C Light Source
Germicidal Lamps 40W UV-C Lamps (18)
Output
254 nm UV-C Light
Chamber Components
Controller
See Touchscreen Display section.
Emergency Switch
Located on both sides and inside the machine, buttons turn
OFF chamber in case of emergency (fire, smoke, etc.).
Tray / Reflective Liner The tray and reflective liner rest inside the bottom of the
chamber and are removable for easy cleaning.
Windows
The tempered glass windows are opaque to UV-C.
Germicidal Lamps
There are eighteen (18) 40W germicidal UV-C lamps.
Magnet
Securely holds doors in closed position while running.
Door Gasket
Seals chamber during operation to keep light from escaping.
3