OPERATION
12
TECHNICAL OPERATING PROPERTIES
BATTERIES
Nominal voltage :
7,2 Vdc
Capacity :
0,5 Ah
Operating time :
>
40 hours, approximately
MOTOR
Voltage applied :
5.5 Vdc ± 0.5 Vdc
Rotational speed :
7000 RPM
(6800 RPM with the instrument open)
Power consumption :
15 mA ± 2 mA at 7000 RPM
FLOW RATES
CIP 10 R :
10 l/min
(at 7000 RPM or 6800 RPM without the housing)
CIP 10 T :
7 l/min
CIP 10 I :
10 l/min
POLYURETHANE FOAMS
Three different polyurethane foams are used:
two in the CIP 10 R selector and one in the rotating cup that is common to
all the models.
These are their properties:
Ext. Diameter Int. Diameter Thickness
Grade
(mm)
(mm)
(mm)
(*)
Selector
Impactor
23
7
5
45
CIP 10 R
}
Selective foam
20
8
24
45
Rotating Cup
36
15
10
60
(*) grade: Number of pores per linear inch