3.11 DC measurement input
Depending on the configuration in the controlling software or device, the DC measurement input can
measure DC voltages or DC currents.
Characteristic
Specification
Current mode
Measurement range
0 ... ±1 mA
0 ... ±20 mA
Typical input resistance
15 Ω
Voltage mode
Measurement ranges
0 ... ±10 V
0 ... ±1 V
0 ... ±100 mV
0 ... ±10 mV
Typical input resistance
1 MΩ
Accuracy in voltage mode
Range
Accuracy
±(% of r % of range)
Temperature coefficient
±(% of r % of range)/°C
beyond 23 °C ± 5 °C
24 hours
1 year
2 years
10 V
0.01 + 0.01
0.03 + 0.01
0.04 + 0.01
0.001 + 0.0003
1 V
0.01 + 0.01
0.03 + 0.02
0.04 + 0.02
0.001 + 0.0003
100 mV
0.01 + 0.02
0.05 + 0.04
0.06 + 0.04
0.001 + 0.0010
10 mV
0.01 + 0.06
0.06 + 0.16
0.07 + 0.16
0.001 + 0.0020
Accuracy in current mode
Range
Accuracy
±(% of r % of range)
Temperature coefficient
±(% of r % of range)/°C
beyond 23 °C ± 5 °C
24 hours
1 year
2 years
20 mA
0.01 + 0.01
0.04 + 0.01
0.05 + 0.02
0.001 + 0.0003
1 mA
Technical data
OMICRON
37
Содержание CMC 430
Страница 1: ...CMC 430 User Manual...
Страница 53: ...ENU 1169 05 02...