2.11 Beeper
A beeper indicates the output of voltages and/or currents. The behavior of the beeper can be
configured in the
CMC 430
Web Interface (accessing the Web Interface → Getting Started manual of
the controlling software).
Beeper mode
Description
Beeper enabled
The beeper is enabled by default. Two modes can be selected:
•
One short beep (0.1 s) at the beginning of the voltage/current output
on the outputs VOLTAGE OUTPUT 1 ... 6,
CURRENT OUTPUT 1 ... 3 or AUX DC OUTPUT
•
Intermittent beep (0.1 s on/2 s off) during the output of voltages/
currents on the outputs VOLTAGE OUTPUT 1 ... 6 and/or
CURRENT OUTPUT 1 ... 3; short beep at the beginning of the
voltage output on output AUX DC OUTPUT
Overload indication:
Intermittent beep (0.1 s on/0.5 s off) in case of an overload on one of
the outputs VOLTAGE OUTPUT 1 ... 6, CURRENT OUTPUT 1 ... 3 or
AUX DC OUTPUT
Beeper disabled
The beeper can be disabled temporarily.
The output AUX DC OUTPUT can be programmed to automatically output a configured
voltage after the start-up of the
CMC 430
. If so, the
CMC 430
issues an intermittent beep
(0.1 s on/0.5 s off) for a period of 20 s after start-up to indicate that voltage will be applied at
this output upon expiry of this period. This warning beep cannot be configured or disabled!
→
If you reduce the voltage setting for the AUX DC OUTPUT from a high value of, for example
220 V, to a low value of, for example 24 V, and you have enabled the beeper in the
CMC 430
Web Interface, you will hear an intermittent beep while the voltage reduces from the high to
the low value.
CMC 430 User Manual
16
OMICRON