2.7 CURRENT OUTPUT section
Two ranges can be configured for the three current outputs: 1.25 A or 12.5 A.
The current outputs are galvanically separated from all other connections of the
CMC 430
.
The current outputs are protected against open-circuit and overload conditions.
The current amplifier is additionally protected against overtemperature.
An LED in the upper left corner and the internal beeper indicate the occurrence
of overload conditions and that current is issued at one or more current outputs
(beeper configuration →
Overload conditions are also indicated in the controlling software or device.
LED state
Description
Red
Current amplifier is active
Red blinking
Overload (burden too high)
OFF
Current amplifier is inactive
NOTICE
Equipment damage possible.
Infeed of currents from external sources may damage the current outputs.
Only in the non-operative state, the current outputs are protected by short-
circuit relays against external infeed.
Technical data →
The current outputs are also available on the generator combination socket →
Device overview