Phase accuracy
Reference
Frequency
Accuracy
24 hours
2 years
V
1
50/60 Hz
0.005°
0.02°
10 Hz ... 100 Hz
-
0.03°
Phasor (UTC)
50/60 Hz
0.005°
0.02°
Note:
Specifications for voltage amplifier channel V
1
are valid with no load on the voltage amplifier.
3.4.3 Total harmonic distortion and noise (THD+N)
The output signal quality depends on the design of the amplifier and its power supplies. The THD+N
values are not subjected by a long-term drift. The following guaranteed THD+N points are verified
during production using a measurement bandwidth of 20 kHz.
Range
Frequency
Current
THD+N (2 years)
1.25 A
55 Hz
0.25 A
< 0.25 %
0.5 A
< 0.07 %
1.25 A
< 0.07 %
12.5 A
55 Hz
1 A
< 0.10 %
5 A
< 0.05 %
12.5 A
< 0.05 %
Note:
These values are also guaranteed between 10 Hz and 100 Hz by design.