MASTER ABSENT
The switching on order of the scales was not respected by switching on a SLAVE scale before its
MASTER or SUBMASTER or a SUBMASTER scale before the MASTER. The message may also appear
during the operation of a SLAVE or SUBMASTER scale in the event that the connection with your
MASTER fails. Check the correct functioning of the MASTER scale, any other SLAVE scales and then
try to switch the scale off and on again.
ATTENTION
: this error could mean that the MASTER or SUBMASTER scale of a bank has problems
with the management of the SLAVES connected to them. In this case, in order to restore operation
of the concerned bench as soon as possible, it is necessary to "create" a new MASTER or
SUBMASTER scale as follows:
-
turn off all the scales on the bench
-
switch on a SLAVE and when the date appears press the key"
ENTER
"
to enable it
- when the scale display shows "- - TEST - -" press the "
C.
”<23>: the scale will automatically
become the MASTER or SUBMASTER of the bank
-
switch the other scales on the bench back on
If the message reappears, call the After-Sales Service.
MEMORY SATURE
The balance of the balance is full. Reset the totals and if the message reappears call the Assistance
Service to install any memory expansion.
Freestyle Compact user manual
Rev.1
Page 35 of 92
Содержание FREESTYLE COMPACT CS
Страница 1: ...USER MANUAL FREESTYLE COMPACT Translated from Italian to English www onlinedoctranslator com...
Страница 2: ...VIEW AND DIMENSIONS MODEL CS Freestyle Compact user manual Rev 1 Page 2 of 92...
Страница 3: ...MODEL CS P wall unit CS PA MODEL Freestyle Compact user manual Rev 1 Page 3 of 92...
Страница 4: ...PLC MODEL Freestyle Compact user manual Rev 1 Page 4 of 92...
Страница 5: ...MODEL PL Freestyle Compact user manual Rev 1 Page 5 of 92...