Thermometer Operating Instructions

CAUTION
1. WHEN DEVICE IS IN USE, DO NOT LOOK DIRECTLY INTO THE LASER BEAM –PERMANENT EYE DAMAGE

MAY RESULT.

2. USE EXTREME CAUTION WHEN OPERATING THE LASER.
3. NEVER POINT THE DEVICE TOWARDS ANYONE’S EYES.
4. KEEP OUT OF REACH OF ALL CHILDREN.

STORAGE & CLEANING The thermometer should be stored at room temperature between –20 to +65

o

C (-4~149°F).

The sensor lens is the most delicate part of the thermometer. The lens should be kept clean at all times, care should be taken when
cleaning the lens using only a soft cloth or cotton swab with water or medical alcohol. Allowing the lens to fully dry before using
the thermometer. Do not submerge any part of the thermometer.

LCD ERROR MESSAGES

The thermometer incorporates visual diagnostic messages as follows:

‘Hi’ or ’Lo’ is displayed when the temperature being measured is outside of the settings of
HAL and LAL.

_

_ _

_

‘Er2’ is displayed when the thermometer is exposed to rapid changes in the ambient

temperature. ‘Er3’ is displayed when the ambient temperature exceeds 0

o

C (32°F) or +50

o

C

(122°F). The thermometer should be allowed plenty of time (minimum 30 minutes) to
stabilize to the working/room temperature.

_

_

Error 5~9

, for all other error messages it is necessary to reset the thermometer. To reset it, wait for

auto power off, remove the battery and wait for a minimum of one minute, reinsert the battery and
turn on. If the error message remains please contact the Service Department for further assistance.

_

__

_

‘Hi’ or ’Lo’ is displayed when the temperature being measured is outside of the measurement range.

BATTERIES

The thermometer incorporates visual low battery indication as follows:

‘Battery OK’: measurements

 are possible

‘Battery Low’: battery needs to be

replaced, measurements are still possible

‘Battery Exhausted’: measurements are

not possible

When the ‘Low Battery’ icon indicates the battery is low, the battery should be replaced immediately with AAA, 1.5V batteries.

Please note: It is important to turn the instrument off before replacing the battery otherwise the thermometer may malfunction.

 Dispose of used battery promptly and keep away from children.

SPECIFICATION

Item

Non-contact Infrared Scan function

Measurement Range

-60 to +860

o

C (-76 to +1580

o

F)

Operating Range

0 to +50°C (32 to +122°F)

Accuracy
(Tobj=15~35°C, Tamb=25°C)

±1.0°C (1.8°F)

Accuracy
(Tobj=-33~860°C,
Tamb=23±3°C)

±2% of reading or 2°C (4°F) whichever is greater

Emissivity Range

0.95 default – adjustable 0.1 to 1 step .01

Resolution (-9.9~199.9°C)

0.1°C/0.1°F

Response Time (90%)

1sec

Distance:Spot

30:1

Battery Life

Typ.180, min 140 hours continuous use (Alkaline, without Laser and Back Light.)

Dimensions

47.0 x 154.9 x 193.1mm(1.8_6.0_7.6 inch)

Weight

344.8 grams(12.16 oz) including batteries (AAA*2pcs)

Note: Under the electromagnetic field of 3V/m from 200 to 600 MHz, the maximum error is 10

o

C (18

o

F).

EMC/RFI:

Readings may be affected if the unit is operated within radio frequency electromagnetic field strength of

approximately 3 volts per meter, but the performance of the instrument will not be permanently affected.

Содержание OS423-LS

Страница 1: ...omega com e mail info omega com For latest product manuals omegamanual info User sGuide OS423 LS Non contact Infrared Thermometer with White LED Flashlight Shop online at MADE IN TAIWAN ...

Страница 2: ...y _ 4 or Down key _ 2 key to change the High Alarm HAL or Lo Alarm LAL then press Meas key _ 6 to confirm it For example When the reading 26 9 o C LAL 27 o C the Low icon will flash and you will hear a beep sound The thermometer will automatically shut off if left idle for more than 60sec ADD VALUE Press Up key _ 4 for LOCK mode ON OFF The lock mode is particularly useful for continuous monitoring...

Страница 3: ...rvice Department for further assistance _ __ _ Hi or Lo is displayed when the temperature being measured is outside of the measurement range BATTERIES The thermometer incorporates visual low battery indication as follows Battery OK measurements are possible Battery Low battery needs to be replaced measurements are still possible Battery Exhausted measurements are not possible When the Low Battery ...

Страница 4: ...y nuclear installation or activity or 2 in medical applications or used on humans Should any Product s be used in or with any nuclear installation or activity medical application used on humans or misused in any way OMEGA assumes no responsibility as set forth in our basic WARRANTY DISCLAIMER language and additionally purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or dam...

Отзывы: