13
table below. Failure to maintain these minimum distances would
invalidate the EN61010 safety approval.
It is essential that following the installation of the system, and prior to
powering the system up that it is tested to determine that the correct
level of isolation is present to protect the user and other equipment
against the hazards of electric shock and fire.
An explanation of creepage and clearance is given in the following
diagram.
Clearance (mm)
Creepage (mm)
Voltage used or generated
by the other assemblies
0.2
1.2
Up to 50V
rms
or V
DC
0.2
1.4
Up to 100V
rms
or V
DC
0.5
1.6
Up to 150V
rms
or V
DC
1.5
3.0
Up to 300V
rms
or V
DC
Above 300V
rms
or V
DC
For a host or other assemblies fitted in the system, using or
generating voltages greater than 300V(rms or DC),advice from
a competent engineer must be obtained before installation of
the relevant equipment.