DP6060 Dual-Input Process Meter
Instruction Manual
71
Display Parameter Action/Setting
Divide
Divide
Channel C = ((A/B)+P)*F
Hi-ab
Max of A or B
C = ((High value of channel A or B)+P)*F
Lo-ab
Min of A or B
C = ((Low value of channel A or B)+P)*F
Drauw
Draw
Channel C = ((A/B)-1)*F
uwAvg
Weighted avg.
Channel C = ((B-A)*F)+A
ratio
Ratio
Channel C = (A/B)*F
Concen
Concentration
Channel C = (A/(A+B))*F
Const
Constant
Constant used in channel C math
adder
Adder
Addition constant used in channel C math
calculations (P)
factor
Factor
Multiplication constant used in channel C
math calculations (F)
CutofF
Cutoff
Set low-flow cutoff
Ch-A
Channel A
Set low-flow cutoff for Channel A
Ch-b
Channel B
Set low-flow cutoff for Channel B
AoutPr
Analog output
programming
Program analog output parameters
AOut 1
*
Analog output 1
Program analog output 1 (*1-3)
parameters
Source
Source
Select source for the 4-20 mA output
break
Loop break
Set relay condition if loop break detected
O-rang
Overrange
Program mA output for display overrange
u-rang
Underrange
Program mA output for display underrange
nmAx
Maximum
Program maximum mA output allowed
nmin
Minimum
Program minimum mA output allowed
Calib
Calibrate
Calibrate 4-20 mA output (internal
reference source used for scaling the
output)
4 nmA
4 mA output
Enter mA output value read by milliamp
meter with at least 0.001 mA resolution
20 nmA
20 mA output
Enter mA output value read by milliamp
meter with at least 0.001 mA resolution
User
User I/O
Assign function keys and digital I/O
F1
*
F1* function key
Assign F1 function key (*F1/F2/F3)
F4
F4 function
Assign F4 function (digital input)
dI I
Digital input 1
Assign digital input 1 – 8, if expansion
modules are connected
dO 1
Digital output 1
Assign digital output 1 – 8, if expansion
modules are connected
ICal
Internal
calibration
Enter internal calibration (used for
recalibrating the meter with a calibrated
signal source)
Ch-A
Channel A
Perform calibration on channel A
Ch-b
Channel B
Perform calibration on channel B