- 6 -

Do not attempt re-adjustment.   

  The highest position of the condenser has been factory adjusted. Do not attempt to 

re-adjustment.  

3.7  Adjusting iris aperture diaphragm   

Move the iris diaphragm lever left or right to adjust the 
aperture size.   

Note:

 

The iris diaphragm is designed to adjust the aperture 

size, not to adjust brightness. Generally, opening the 
diaphragm to 70-80% of the N.A. value of the 
respective objective will provide an image of 
acceptable quality. If you want to observe the image of 
the iris diaphragm, remove one eyepiece and look 
through the tube. You will see a dark circle encroaching 
on the bottom of the tube.   

3.8

 

Adjusting focus knob tension

   

The tightness of the focus knob tension has been 
pre-set at the factory. If the mechanical stage 
drops by itself, rotate the tension adjustment ring 
located inside the focus knob on the power switch 
side until the tension is in maintained. Tension 
Ring  

 

 
 
 
 
 
 
 

 
 
 
 
 
 
 
 
 
 
 
 

4

 

Specifications 

Tension Ring 

Содержание M825

Страница 1: ...User Manual Binocular Compound Microscope Model M825 MicroscopeNet com...

Страница 2: ...Table of Contents i Caution 1 ii Care and Maintenance 2 1 Component Illustration 3 2 Installation 4 3 Operation 5 4 Specifications 7 5 Troubleshooting Guide 8...

Страница 3: ...ure or humidity and dusty environments Ensure that the microscope is located on a smooth level and firm surface 4 If any specimen solutions or other liquids splash onto the stage objective or any othe...

Страница 4: ...he optical surface use a soft cloth or cotton swab dampened with a lens cleaning solution available at camera stores All optical lenses should be swabbed using a circular motion A small amount of abso...

Страница 5: ...ece Tube 3 Diopter Ring 4 Viewing Head 5 Slide Holder 6 Stage 7 Condenser 8 Light Collector 9 Set Screw 10 Nosepiece 11 Objective 12 Stand 13 Coarse Focus Knob 14 Fine Focus Knob 15 Translation Knobs...

Страница 6: ...2 3 Installation of the objectives 1 Adjust the coarse focus knob until the mechanical stage is at its lowest position 2 Install the 4x objective into the nosepiece Then in a clock wise direction rot...

Страница 7: ...tive turn the 100X objective in position and rotate the focus stop lever clock wise all the way to the end so that the 100X objective will not contact the specimen while the stage is adjusted to its h...

Страница 8: ...m to 70 80 of the N A value of the respective objective will provide an image of acceptable quality If you want to observe the image of the iris diaphragm remove one eyepiece and look through the tube...

Страница 9: ...ves Achromatic 4X 10X 40X spring 100X spring oil Condenser Abbe NA 1 25 w iris aperture diaphragm and filter holder Rack and pinion adjustment Focus Mechanism Coaxial coarse and fine focusing knobs on...

Страница 10: ...d to the slide Attach a 0 17mm slide cover Slide cover is too thick or thin Use a slide cover of the appropriate thickness 0 17mm Slide may be upside down Turn slide over so the cover glass faces up P...

Страница 11: ...tensity control dial Aperture diaphragm closed too far Open to the proper setting Condenser position too low Position the condenser upward IMAGE PROBLEMS Problem Cause Solution Image will not focus wi...

Отзывы: