background image

 

 

www.microscopenet.com

 

 

- 2 -

ii.

 

Care and Maintenance   

1.  Do not attempt to disassemble any component including eyepieces, objectives or 

focusing assembly.   

2.  Keep the instrument clean; remove dirt and debris regularly. Accumulated dirt on metal 

surfaces should be cleaned with a damp cloth. More persistent dirt should be removed 
using a mild soap solution. Do not use organic solvents for cleansing.   

3.  The outer surface of the optics should be inspected and cleaned periodically using an 

air stream from an air bulb. If dirt remains on the optical surface, use a soft cloth or 
cotton swab dampened with a lens cleaning solution (available at camera stores). All 
optical lenses should be swabbed using a circular motion. A small amount of absorbent 
cotton wound on the end of a tapered stick makes a useful tool for cleaning recessed 
optical surfaces. Avoid using an excessive amount of solvents as this may cause 
problems with optical coatings or cemented optics or the flowing solvent may pick up 
grease making cleaning more difficult. Oil immersion objectives should be cleaned 
immediately after use by removing the oil with lens tissue or a clean, soft cloth.   

4.  Store the instrument in a cool, dry environment. Cover the microscope with the dust 

cover when not in use.   

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Содержание M825

Страница 1: ...User Manual Binocular Compound Microscope Model M825 MicroscopeNet com...

Страница 2: ...Table of Contents i Caution 1 ii Care and Maintenance 2 1 Component Illustration 3 2 Installation 4 3 Operation 5 4 Specifications 7 5 Troubleshooting Guide 8...

Страница 3: ...ure or humidity and dusty environments Ensure that the microscope is located on a smooth level and firm surface 4 If any specimen solutions or other liquids splash onto the stage objective or any othe...

Страница 4: ...he optical surface use a soft cloth or cotton swab dampened with a lens cleaning solution available at camera stores All optical lenses should be swabbed using a circular motion A small amount of abso...

Страница 5: ...ece Tube 3 Diopter Ring 4 Viewing Head 5 Slide Holder 6 Stage 7 Condenser 8 Light Collector 9 Set Screw 10 Nosepiece 11 Objective 12 Stand 13 Coarse Focus Knob 14 Fine Focus Knob 15 Translation Knobs...

Страница 6: ...2 3 Installation of the objectives 1 Adjust the coarse focus knob until the mechanical stage is at its lowest position 2 Install the 4x objective into the nosepiece Then in a clock wise direction rot...

Страница 7: ...tive turn the 100X objective in position and rotate the focus stop lever clock wise all the way to the end so that the 100X objective will not contact the specimen while the stage is adjusted to its h...

Страница 8: ...m to 70 80 of the N A value of the respective objective will provide an image of acceptable quality If you want to observe the image of the iris diaphragm remove one eyepiece and look through the tube...

Страница 9: ...ves Achromatic 4X 10X 40X spring 100X spring oil Condenser Abbe NA 1 25 w iris aperture diaphragm and filter holder Rack and pinion adjustment Focus Mechanism Coaxial coarse and fine focusing knobs on...

Страница 10: ...d to the slide Attach a 0 17mm slide cover Slide cover is too thick or thin Use a slide cover of the appropriate thickness 0 17mm Slide may be upside down Turn slide over so the cover glass faces up P...

Страница 11: ...tensity control dial Aperture diaphragm closed too far Open to the proper setting Condenser position too low Position the condenser upward IMAGE PROBLEMS Problem Cause Solution Image will not focus wi...

Отзывы: