24
Chapter 3 Preparation and Inspection
OSF TYPE V60 OPERATION MANUAL
Attaching the Air/Water Valve
Apply a small amount of lubricant to the Seals. Attach the Air/Water Valve to the
Air/Water Cylinder of the endoscope (see Figure 3.9). Confirm that the valve fits
properly without bulging of the Skirt.
Figure 3.9
The Air/Water Valve may ‘stick’ at first, but it should operate
smoothly after it is depressed a few times.
Attaching the Biopsy Valve
Attach the Biopsy Valve to the Instrument Channel Port of the endoscope (see
Figure 3.10). Confirm that the Cap fits properly without bulging.
Figure 3.10
Skirt
Air/Water Valve
(MB-196)
Suction Cylinder
Air/Water Cylinder
Assembled
Suction Valve
(MB-197)
Biopsy Valve
Assembled
Instrument
Channel Port