Chapter 3 Preparation and Inspection
19
OSF TYPE V60 OPERATION MANUAL
4.
Holding the Insertion Tube gently with one hand, carefully run your fingertips
over the entire length of the Insertion Tube in both directions (see Figure
3.2). Confirm that no objects or metallic wire protrude from the Insertion
Tube. Also confirm that the Insertion Tube is not abnormally rigid.
Figure 3.2
5.
Using both hands, bend the Insertion Tube of the endoscope into a
semicircle. Then, moving your hands as shown by the arrows in Figure 3.3,
confirm that the entire Insertion Tube can be smoothly bent to form a
semicircle and that the Insertion Tube is pliable.
Figure 3.3
6.
Gently hold the midpoint of the Bending Section and a point 20 cm from the
Distal End. Push and pull gently to confirm that the junction between the
Bending Section and the Insertion Tube is not loose.
7.
Inspect the objective lens at the Distal End of the endoscope’s Insertion
Tube for scratching, cracks, stains, gaps around the lens or other
irregularities.
8.
Inspect the air/water nozzle at the Distal End of the endoscope’s Insertion
Tube for abnormal swelling, bulges, dents or other irregularities.