Chapter 4 Operation
37
OSF TYPE V60 OPERATION MANUAL
• When inserting or withdrawing an Endo-Therapy Accessory,
confirm that its Distal End is closed or completely retracted
into the sheath. Slowly insert or withdraw the Endo-Therapy
Accessory straight into/from the slit of the Biopsy Valve.
Otherwise, the biopsy valve may be damaged and pieces of it
could fall off.
• If the insertion or withdrawal of Endo-Therapy Accessories is
difficult, straighten the Bending Section as much as possible
without losing the endoscopic image. Inserting or
withdrawing Endo-Therapy Accessories with excessive force
may damage the Instrument Channel or Endo-Therapy
Accessories, cause some parts to fall off and/or cause patient
injury.
• If an Endo-Therapy Accessory cannot be withdrawn from the
endoscope, close the tip of the Endo-Therapy Accessory or
retract the tip of the Endo-Therapy Accessory into its sheath
and slowly withdraw the endoscope while observing the
endoscopic image.
• If the Distal End of an Endo-Therapy Accessory is not visible
in the endoscopic image, do not open the Distal End or
extend the needle of the instrument. This could cause patient
injury, bleeding, perforation and/or equipment damage.
Insertion of Endo-Therapy Accessories Into the Endoscope
• Do not insert Endo-Therapy Accessories forcibly or abruptly.
Otherwise, the Endo-Therapy Accessory may extend from
the Distal End of the endoscope abruptly, which could cause
patient injury, bleeding and/or perforation.
• When using the Endo-Therapy Accessory with the cap of the
Biopsy Valve detached, it is easier to insert the accessory.
But, as a result, it can reduce efficacy of the endoscope’s
suction system, and may leak or spray patient debris or
fluids, posing an infection control risk. When not using the
Endo-Therapy Accessory, attach the cap to the main body of
the Biopsy Valve.
• When the cap of the Biopsy Valve is detached, it may cause
patient debris or fluids to leak or spray from the endoscope,
posing an infection control risk. When the valve is uncapped,
place a piece of sterile gauze over it to prevent leakage.