18
Chapter 3 Preparation and Inspection
OSF TYPE V60 OPERATION MANUAL
Figure 3.1
3.2
Preparation and Inspection of the Endoscope
Clean and disinfect or sterilize the endoscope as described in the endoscope’s
companion manual, the “OLYMPUS OSF TYPE V60 ENDOSCOPE
REPROCESSING MANUAL”.
Inspection of the Endoscope
1.
Inspect the control section and the endoscope connector for excessive
scratching, deformation, loose parts or other irregularities.
2.
Inspect the Boot and the Insertion Tube near the Boot for bends, twists or
other irregularities.
3.
Inspect the external surface of the entire Insertion Tube including the
Bending Section and the Distal End for dents, bulges, swelling, scratching,
holes, sagging, transformation, bends, adhesion of foreign bodies, dropout
of parts, any protruding objects or other irregularities.
• Paper Towels
• Trays
• Personal Protective Equipment
• Lint-free Cloths
Endo-Therapy Accessory
Biopsy Valve
Endoscope
Suction Pump
Video System Center
Display Monitor
Water Container