USG-400
ISSUE1 6-3
Safety
check
1-4
Patient leakage current confirmation
○
Criteria
The patient leakage current must not be the value at the table 1-5.
○
Inspection procedures
(1) Turn off the power switch of USG-400.
(2) Connect the Leakage current measuring jig
(SS0499) to USG-400, and then connect the
SONICBEAT Transducer with the SONICBEAT instrument (SB-0545PC or SB-0545IC)
attached to it to the opposite side.
USG-400
Leakage current measuring