USG-400
ISSUE1
3-3
Installation and Connection
2 Installation of the ultrasonic generator
CAUTION
Heed the following cautions when installing the ultrasonic generator in
a position other than on the energy cart shown in the “System chart” in
the Product Outline.
−
Install the ultrasonic generator on a level, stable platform.
If the ultrasonic generator slips from or topples down from an u
nstable installation position, the patient, surgeon and/or surgical
staff may be injured. Furthermore, equipment damage may
occur.
−
Confirm that the installation position has enough strength and
size for supporting the ultrasonic generator. If it is not strong
enough, the ultrasonic generator may fall and be damaged.
Be careful when the trolley (cart) containing the ultrasonic
generator and other equipment are transported on a non-level
surface or across an obstacle cable. If the trolley (cart) is
transported over an obstacle, the ultrasonic generator and other
equipment may tip over possibly resulting in equipment failure.
Do not install the ultrasonic generator on its side or upside down.
Otherwise, a failure or malfunction may occur.
Do not install the ultrasonic generator in close contact with a
wall or other equipment in a manner that would block the
ventilation opening. Otherwise, the generator may overheat.
Be careful not to pinch a hand when installing the ultrasonic
generator.
Avoid applying excessive force on the socket connections.
Otherwise, wire disconnection or other failure may occur.
Always use the ultrasonic generator in compliance with the
“Operating environment” specified in the Product Outline.
Otherwise, the ultrasonic generator may fail.
NOTE
Keep the instruction manual in an easily accessible place near
the ultrasonic generator.
Before connecting an accessory to the ultrasonic generator, read
And understand the instruction manual for the accessory
thoroughly and confirm the compatibility of the ultrasonic
generator.
1.
Checking the
operationg conditions
Confirm that the operating conditions comply with the information provided in
“Danger, warnings, and cautions.
2.
Installing the trolley (cart)
When using the energy cart TC-E400, prepare the cart by following its instruction
manual.
3.
Installing the ultrasonic generator
Place the ultrasonic generator on a stable trolley (cart) or a level, stable platform.
Use of the energy cart TC-E400 is highly recommended.
Содержание USG-400
Страница 1: ...ISSUE 1 2011 11 MAINTENANCE MANUAL USG 400...
Страница 2: ......
Страница 6: ......
Страница 10: ...USG 400 Product Outline 1 4 ISSUE1 4 System chart...
Страница 11: ...USG 400 ISSUE1 2 1 Specifications Chapter 2 Specifications 1 Specifications 2 2...
Страница 14: ......
Страница 34: ......
Страница 38: ......
Страница 43: ...USG 400 ISSUE1 5 5 Inspection Figure 5 3 Figure 5 4...
Страница 50: ...USG 400 Inspection 5 12 ISSUE1 Figure 5 12...
Страница 59: ...USG 400 ISSUE1 6 1 Safety check Chapter 6 Safety check 1 Check procedures 6 2 2 Check Card 6 6...
Страница 65: ...USG 400 ISSUE1 7 1 Block Diagram Chapter 7 Block Diagram 1 Block Diagram 7 2...
Страница 89: ......