USG-400
Inspection 5-16
ISSUE1
Figure 5.16
3.
Activation
Press and hold the SEAL button (blue) of the handswitch or the SEAL pedal
(right pedal: blue) of the footswitch for the high-frequency (RF bipolar)
output.
4.
Inspectiong the output
Confirm the following items during the SEAL mode output:
- The activation “All screen” or “Set screen” is displayed on the touch-screen
of the ultrasonic generator, no error window (see Figure 5.17).
- The output tone sounds from the compatible electrosurgical generator.
Figure 5.17