Chapter 7 Troubleshooting
61
ELECTROSURGICAL UNIT PSD-30
When the FL code is displayed
Indication
of the cut
output
setting
indicator
Indication
of the
coagulation
output
setting
indicator
Possible cause
Solution
The foot switch is
short-circuited.
Replace the foot switch with
a new one.
• The cut mode is not
selected.
• The coagulation mode
is not selected.
Select the cut mode or
coagulation mode as
described in Section 5.2,
“Selection of cut mode” or
5.3, “Selection of
coagulation mode”.
The output level is set to
0 W.
Set the output level to a
value above 0 W.
The electrosurgical unit is
malfunctioning.
Immediately turn the
electrosurgical unit OFF and
contact Olympus.