Chapter 5 Operation
ELECTROSURGICAL UNIT PSD-30
−
When cutting a mucous membrane, set the output level
and output time to the lowest necessary limits at which
safe cutting is possible. Otherwise, perforation may result.
It is recommended that you perform basic experiments
before using the electrosurgical unit.
−
When performing coagulation, set the output level and
output time to the lowest necessary limits at which safe
coagulation is possible. Otherwise, perforation may result.
It is recommended that you perform basic experiments
before using the electrosurgical unit.
• Precautions for endoscopic papillotomy
−
Set the output level and output time to the lowest
necessary limits at which safe cutting is possible.
Otherwise, perforation and/or bleeding may result.
−
Confirm that electricity is supplied to the accessory when
making an incision. Incision without electricity may result
in perforation and/or bleeding.
−
Use only the cutting mode. Using the coagulation mode
may result in perforation and/or damage to the accessory.
• Precautions for hot biopsy
−
Set the output level and output time to the lowest
necessary limits at which safe cutting is possible.
Otherwise, perforation and/or bleeding may result.
−
Do not output while the metal portion of the hot biopsy
forceps is in contact with a tissue in other part that the
position where it grasps the tissue. Otherwise, perforation
and/or burns to the tissue in contact may result.
−
Do not apply the cup of the hot biopsy forceps to the
tissue but pull the tissue sufficiently to avoid perforation.
−
If the biopsy tissue cannot be cut, check the tissue contact
condition and output setting again. Note that setting an
excessive output level may result in perforation and/or
bleeding.
5.1
Turn the power ON
Confirm that the accessories for the procedure have been connected correctly
as described in Chapter 3, “Installation and Connection” and turn the
electrosurgical unit ON.