Chapter 5 Operation
55
ELECTROSURGICAL UNIT PSD-30
5.8
Adjusting the volume
The output tone volume can be adjusted in four steps. Since
the output tones play an important role of noticing the output,
do not lower the volume down to the inaudible area. If the
output tone is inaudible, high-frequency output may not be
detected by the user. This could cause patient injury.
1.
Put the electrosurgical unit in the stand-by mode as described in Section
5.7, “Stand-by mode”.
2.
Confirm that the numbers 1 to 4 are displayed on the coagulation output
setting indicator. The larger the number, the larger the output tone volume.
3.
Press the “
” segment of the output control button on the COAG. side to
increase the displayed number (see Figure 5.13). Press the “
” segment to
decrease the displayed number.
Figure 5.13
4.
After completing the adjustment, cancel the stand-by mode as described in
Section 5.7, “Stand-by mode”.
• The output tone volume can be adjusted only in the stand-by
mode.
• The alarm tone volume cannot be adjusted.