Chapter 4 Inspection
ELECTROSURGICAL UNIT PSD-30
Figure 4.1
Figure 4.2
If the power fails to come on
If the power fails to come on, inspect the system as follows:
1.
Confirm that the power cord is connected securely to a wall mains outlet and
the AC power inlet on the electrosurgical unit.
Turn the electrosurgical unit OFF and remove the power cord
from the AC power inlet on the electrosurgical unit before
removing the fuse box. Otherwise, fire or electric shock may
result.
2.
Turn the electrosurgical unit OFF and disconnect the power cord from the
wall mains outlet.
3.
Remove the fuse box from the electrosurgical unit by squeezing the tabs on
the sides of the fuse box and pulling it straight out (see Figure 4.3).
Power switch