Important Information — Please Read Before Use
ELECTROSURGICAL UNIT PSD-30
• Flammable agents used for cleaning and disinfection must be
allowed to evaporate before the electrosurgical unit is used.
Also ensure that flammable solutions are neither on the
patient’s skin nor in the patient’s body cavity when the
electrosurgical unit is used.
• To prevent patient burns when the electrosurgical unit is used
in conjunction with a non-Olympus electrosurgical unit,
ensure that each unit is supplied with power from wall mains
outlets. Also ensure that each unit is supplied with power
from separate breakers.
• During operation, temporarily unused electrodes should be
stored in an electrically insulated container. Unused
electrodes should never be placed on the patient otherwise it
may cause operator and/or patient burns.
• The electrosurgical unit should not be used for performing
circumcisions.
• To prevent patient burns, never use cords with scratches or
cracks.
• Studies have shown that smoke generated during
electrosurgical procedures can be irritating and potentially
harmful to surgical personnel. These studies recommend the
use of surgical masks and adequate ventilation of smoke by
the use of surgical smoke evacuators or other means.