– 9 –
11
Fix the drape with 3 straps. Beginning from the
position closer to the microscope section,
attach 3 straps in the same way as above.
12
Press the scope switch C on the microscope section, move the arm, and
confirm that its movement is not hindered. Adjust the swelling of the
drape near the strap when the arm movement is not smooth or when the
switch is unintentionally pushed by the drape. If it still does not resolve
please remove the strap and repeat from Step 11.
13
Check that there is no breakage or adhesion of foreign matter, wetting due to water, etc. If there is an
abnormity, replace with new drape.
14
Check the image of observation monitor is no abnormality.
Fix on the tip side of the
black part.
Black part
Cut out the straps along
the perforations.
Peel off the release paper and
attach it onto the same strap.
How to use the strap