– 4 –
2.4 Inspection of the power ON
2.5 Inspection of the illumination light
2.6 Inspection of the observation monitor
2.7 Inspection of the image
1
Check that the ventilation grilles on the front and rear
of the base of this instrument are not stopped by dust,
a foreign object, or cloth.
2
Confirm that the power switch of the LED light source is set to “ (ON)” as described in the instruction manual
for the LED light source.
3
Press the power switch of this instrument to turn it ON. The power indicator
lights green and the illumination lamp of the LED light source lights up.
Confirm that the illumination light is output from the illumination lens of the
microscope section.
Operate the observation monitor by following its instruction manual and confirm that it can display images.
1
Using a white object such as a piece of gauze as the
subject, bring the focus on it and set so that no
clipped white area is observed on the entire image.
2
Keeping this instrument unmovable, tap and hold the
“White balance” button on the touch panel or the
custom switch to which the white balance function is
assigned for about 1 second.
3
The white balance display window shows “Completed”, the message is
displayed on the observation monitor and the white balance icon appears on
the touch panel.
Ventilation
grilles
Power indicator
Power switch
Illumination
lens
White balance
White balance display window
White balance icon