– 11 –
As the measures to be taken until the spare equipment is available, this instrument provides the following remedial
actions.
In case the arm cannot be locked
In case movement by joystick control cannot be stopped
In case focus does not work
In case zoom does not work
1
Flip open the switch cover of this instrument and
press the arm locking switch to lock the arm.
2
When the view field movement stop switch is pressed, movement of the microscope section by pressing the
scope switch C may become heavy.
1
Flip open the switch cover of this instrument and
press the view field movement stop switch to stop the
joystick operation.
2
Press the scope switch C and move the microscope section.
Press the scope switch C and move the microscope section to focus on the observed region.
1
Pull out the manual zoom knob and confirm that the status indication
line (orange) is visible. A message is displayed on the observation
monitor indicating that the manual zoom knob has been pulled out.
2
Turn the manual zoom knob to control zoom.
4 Remedial actions in case of emergency
Switch cover
Arm locking switch
View field movement
stop switch
Switch cover
Manual
zoom knob
Status indication
line (orange)