– 5 –

2.8 Inspection of the focus/zoom functions

2.9 Inspection of the view field movement

4

Place an object having an easy-to judge vertical 
direction view so that it is focused on the center 
of the image displayed on the observation 
monitor. Observe the observation monitor 
images without wearing the 3D eyeglasses and 
confirm that there is no positioning error in the 
up-down direction between the left and right 
images.

5

Observe the images by wearing the 3D 
eyeglasses, close the left and right eyes 
alternately, and confirm that there is no difference in color or brightness between the left and right images.

6

While observing the 3D images, contact the observed object with forceps or a similar tool and confirm that 
there is no peculiarity in the feeling of depth.

7

Confirm that the data display is displayed on the observation monitor.

8

Observe your hand and confirm that the image is displayed 
in normal colors without noise.

1

Tilt the scope switches A and B on the microscope 
section all the way upward or downward and, by 
observing the observation monitor, confirm that the 
focus and zoom functions are activated correctly.

2

If the zoom + of the adaptive AF is set, confirm that 
the auto focus is activated and focus is adjusted after 
completion of the zoom + (UP) operation.

1

Tilt the joystick on the foot switch all the 
way. Observe the observation monitor and 
confirm that the field moves to direction in 
which the joystick is tilted.

2

If the view field movement of the adaptive 
AF is set, confirm that the auto focus is 
activated after completion of the view field 
movement operation.

Observation 
monitor

Images without up-down 
positioning error

Images with up-down 
positioning error

Left image

Right image

Left image

Right image

Data display

Observation 
monitor

Observation monitor

Scope 
switch A

Scope 
switch B

Observation monitor

Joystick

Foot switch

Observation monitor

Содержание ORBEYE OME-V200

Страница 1: ...e and confirm that this instrument is locked to the floor 1 2 Connection of the power cord Connect the power plug of the power cord directly into the hospital grade wall mains outlet Connection of LMD...

Страница 2: ...ME V200 HVO 4000ST MAJ 2310 A Red B Green C Blue D White Playback terminal Release terminal Recording control terminal SDI input terminals A to D 3G SDI quad link combined output 1 or 2 terminal REMOT...

Страница 3: ...w is visible do not use this instrument and contact Olympus 1 Confirm that this instrument and the ancillary equipment to be used are turned OFF 2 Confirm that the electrical contacts of the connector...

Страница 4: ...light is output from the illumination lens of the microscope section Operate the observation monitor by following its instruction manual and confirm that it can display images 1 Using a white object...

Страница 5: ...8 Observe your hand and confirm that the image is displayed in normal colors without noise 1 Tilt the scope switches A and B on the microscope section all the way upward or downward and by observing...

Страница 6: ...nger from the scope switch C and confirm that the arm is locked and the microscope section can be fixed at the current position 4 If the arm free of the adaptive AF is set confirm that the auto focus...

Страница 7: ...ection and rotate the microscope section around rotation axis A shown in a by 180 toward the front in the direction of arrow 1 to the positioning shown in b Then rotate the microscope section around r...

Страница 8: ...grip sheet around the grip section of the microscope section and hold down the slack of the drape Peel off the release paper at the end of the grip sheet and firmly press the attaching bezel and attac...

Страница 9: ...is not smooth or when the switch is unintentionally pushed by the drape If it still does not resolve please remove the strap and repeat from Step 11 13 Check that there is no breakage or adhesion of f...

Страница 10: ...Loading the user preset 1 Tap the Load user preset button 2 Tap the user ID button to be set 3 The selected setting will appear on the user ID window on the touch panel Load user preset User ID User I...

Страница 11: ...come heavy 1 Flip open the switch cover of this instrument and press the view field movement stop switch to stop the joystick operation 2 Press the scope switch C and move the microscope section Press...

Страница 12: ...atient 2 Open the loupes storage section of the base and take out the loupes case 3 Take the loupes out of the loupes case 4 While pulling the loupes lock ring in the direction of the arrow insert the...

Страница 13: ...e off the microscope section 6 Turn the loupes lock ring clockwise to secure the loupes on the microscope section Confirm that the caution line yellow is invisible 7 Hold the loupes adjust the width t...

Страница 14: ...n of the observation monitor Inspection of the image Inspection of the focus zoom functions Inspection of the motorized field movement Inspection of the arm locking unlocking After inspection Store an...

Отзывы: