122
Chapter 6 Measurement procedures (LX only)
Displaying the measurement results
The three most recent measurement results are displayed. The object distance
mark (
) (see "page 125" (3 Displaying the measurement results)) indicates the
object distance of measurement point. This is displayed to the right side of the
measured value.
Re-specifying a measurement point
Select “CLEAR” in the stereo measurement menu to delete measurement points
specified by mistake.
Re-specifying all measurement points
If you want to restart the measurement point specification in an image from the
beginning, select “ALL CLEAR” in the stereo measurement menu.
Matching confirmation and distance display
When matching completes, the relevant corresponding point is displayed in the
right image area. Confirm that it corresponds to the position specified in the left
image area. At this time, the distance from the insertion tube distal end to the
specified measurement point is also displayed as “Z= __ ”. The background
changes yellow when the distance to the measurement point is over 15 mm or
red when it exceeds 30 mm. The distance range up to 15 mm is the
recommended measuring range and 30 mm is the maximum measurable range.
In case auto matching is not smooth
Read "Notes for capturing measurement images" (page 116) and "In case
accuracy cannot be achieved" (page 123).
Recording the measurement result
The image with the measurement result is recorded by pressing the [FRZ/REC]
button for at least two seconds (long push) when the measurement result is
displayed.