98
7.5 Reprocessing the light source and accessories
CLV-190 INSTRUCTION MANUAL
Ch.7
Reprocessing workflow for the light source and accessories
This chapter describes the workflow for reprocessing the light source and accessories.
7.5
Reprocessing the light source and accessories
1
Confirm the appearance of the light source and accessories.
2
If patient debris directly or indirectly adheres to this instrument and accessories, refer
to “
Disinfection” on page 99. Otherwise, be sure to follow the next step.
3
If disinfection is necessary according to national, local, or facility standards, refer to
“
Disinfection” on page 99. Otherwise, refer to “
Cleaning” on page 100.
3. Disinfection is necessary according to
national, local, or facility standards
2. Patient debris directly or indirectly adheres to this instrument
Cleaning
(
→
”
Cleaning”)
Disinfection
(
→
”
Disinfection”)
Yes
Yes
No
No
1. Confirm the appearance of the light source
Содержание EVIS EXERA III
Страница 2: ......
Страница 6: ...iv Contents CLV 190 INSTRUCTION MANUAL...
Страница 20: ...14 Summary of the Equipment Functions CLV 190 INSTRUCTION MANUAL...
Страница 22: ...16 1 1 Checking the package contents list CLV 190 INSTRUCTION MANUAL Ch 1...
Страница 26: ...20 2 2 Front panel CLV 190 INSTRUCTION MANUAL Ch 2 Control panel buttons 13 9 8 7 6 5 14 12 11 10...
Страница 28: ...22 2 2 Front panel CLV 190 INSTRUCTION MANUAL Ch 2 Control panel indicators 15 18 16 26 25 23 17 22 24 20 21 19...
Страница 32: ...26 2 3 Rear and side panels CLV 190 INSTRUCTION MANUAL Ch 2...
Страница 98: ...92 6 4 Lamp usage indicator reset CLV 190 INSTRUCTION MANUAL Ch 6...
Страница 108: ...102 7 8 Transportation CLV 190 INSTRUCTION MANUAL Ch 7...