4.9 Inspection of brightness adjustment
51
CLV-190 INSTRUCTION MANUAL
Ch.4
Table 4.1
Inspection of automatic brightness adjustment
WARNING
When disconnecting the camera head or the video converter from the endoscope
without turning the examination lamp OFF, make sure that the brightness mode
indicator is set to “MANU” and that the light intensity adjustment button is set to the
minimum examination light intensity. If the camera head or the video converter is
disconnected while the brightness mode indicator is set to “AUTO”, the intense light
may cause eye injury.
4.9
Inspection of brightness adjustment
Confirm that the brightness of the examination lamp can be adjusted.
The inspection method differs according to the endoscope to be used. Select either
the “AUTO” or “MANU” mode as shown in Table 4.1.
Brightness
adjustment mode
Connected endoscopes
AUTO
EVIS EXERA III 190 series videoscope
EVIS EXERA 160, EVIS EXERA II 180 series videoscope
EVIS 100, EVIS 130, EVIS 140 series videoscope
OES fiber endoscope used in combination with the video converter
Rigid videoscope
Flexible videoscope
Rigid endoscope used in combination with a camera head and the light guide cable
Fiber endoscope used in combination with a camera head and the light guide cable
MANU
OES fiber endoscope
Rigid endoscope used in combination with the light guide cable
Fiber endoscope used in combination with the light guide cable
Содержание EVIS EXERA III
Страница 2: ......
Страница 6: ...iv Contents CLV 190 INSTRUCTION MANUAL...
Страница 20: ...14 Summary of the Equipment Functions CLV 190 INSTRUCTION MANUAL...
Страница 22: ...16 1 1 Checking the package contents list CLV 190 INSTRUCTION MANUAL Ch 1...
Страница 26: ...20 2 2 Front panel CLV 190 INSTRUCTION MANUAL Ch 2 Control panel buttons 13 9 8 7 6 5 14 12 11 10...
Страница 28: ...22 2 2 Front panel CLV 190 INSTRUCTION MANUAL Ch 2 Control panel indicators 15 18 16 26 25 23 17 22 24 20 21 19...
Страница 32: ...26 2 3 Rear and side panels CLV 190 INSTRUCTION MANUAL Ch 2...
Страница 98: ...92 6 4 Lamp usage indicator reset CLV 190 INSTRUCTION MANUAL Ch 6...
Страница 108: ...102 7 8 Transportation CLV 190 INSTRUCTION MANUAL Ch 7...