104
8.2 Troubleshooting guide
CLV-190 INSTRUCTION MANUAL
Ch.8
Irregularity description
Possible cause
Solution
The endoscope cannot be
connected to the light source.
The endoscope is not compatible with
the light source.
Connect an endoscope that is listed in
the “
System chart” on page 109.
The power fails to turn ON.
The power switch is not turned ON.
Turn the power switch ON.
The power cord is not connected.
Connect it to a wall mains outlet as
described in Section 3.8, “Connection to
the AC mains power supply”.
The lamp cover is not closed firmly or
removed.
Close the lamp cover firmly described in
Section 6.3, “Insertion of the lamp”.
The power switch of the mobile
workstation is OFF.
Turn the power switch of the mobile
workstation ON.
The examination lamp does
not ignite.
The examination lamp has not been
ignited yet.
Press the lamp button.
The examination lamp is not installed.
Install an examination lamp as
described in Section 6.1, “Replacement
of the examination (xenon) lamp”.
The examination lamp is not installed
correctly.
Reinstall the examination lamp as
described in Section 6.1, “Replacement
of the examination (xenon) lamp”.
The examination lamp is broken.
Replace the examination lamp with a
new one as described in Section 6.1,
“Replacement of the examination
(xenon) lamp”.
The examination light is OFF.
Turn OFF the light source and ignite the
examination lamp again.
No light is emitted from the
endoscope.
The examination lamp has not been
ignited yet.
Press the lamp button.
The endoscope or light guide is not
connected to the output socket.
Connect the endoscope to the output
socket securely as described in
Section 4.4, “Connection of an
endoscope”.
The examination lamp does
not ignite automatically when
the light source is turned ON.
The manual ignition mode is selected
to ignite the examination lamp.
Set the lamp ignition mode switch on
the rear panel to “AUTO” as described
in Section 3.5, “Selection of the lamp
ignition mode”.
The examination lamp ignites
automatically when the light
source is turned ON.
The auto ignition mode is selected to
ignite the examination lamp.
Set the lamp ignition mode switch on
the rear panel to “MANU” as described
in Section 3.5, “Selection of the lamp
ignition mode”.
The examination lamp does
not turn OFF even if the lamp
button is pressed.
The lamp button is pressed only for a
short time.
Press and hold the lamp button for
1 second or more.
Содержание EVIS EXERA III
Страница 2: ......
Страница 6: ...iv Contents CLV 190 INSTRUCTION MANUAL...
Страница 20: ...14 Summary of the Equipment Functions CLV 190 INSTRUCTION MANUAL...
Страница 22: ...16 1 1 Checking the package contents list CLV 190 INSTRUCTION MANUAL Ch 1...
Страница 26: ...20 2 2 Front panel CLV 190 INSTRUCTION MANUAL Ch 2 Control panel buttons 13 9 8 7 6 5 14 12 11 10...
Страница 28: ...22 2 2 Front panel CLV 190 INSTRUCTION MANUAL Ch 2 Control panel indicators 15 18 16 26 25 23 17 22 24 20 21 19...
Страница 32: ...26 2 3 Rear and side panels CLV 190 INSTRUCTION MANUAL Ch 2...
Страница 98: ...92 6 4 Lamp usage indicator reset CLV 190 INSTRUCTION MANUAL Ch 6...
Страница 108: ...102 7 8 Transportation CLV 190 INSTRUCTION MANUAL Ch 7...