5.9 Air/water feeding
CLV-190 INSTRUCTION MANUAL
Ch.5
WARNING
Over-insufflating the lumen may cause patient pain, injury, bleeding, gas embolism,
and/or perforation.
5.9
Air/water feeding
1
Confirm that the airflow indicator “ON” lights up.
If not, press the airflow button. The airflow
indicator “ON” lights up to indicate that air is
being fed to the endoscope.
Figure 5.9
2
Press the airflow regulator button to set the
airflow regulator according to the examination
technique and/or patient condition. Each press
of the button changes the airflow regulator
indicators between L (Low), M (Medium), and H
(High).
Figure 5.10
3
Feed water or air as described in the endoscope’s instruction manual.
Airflow indicators
Airflow button
Airflow regulator button
Airflow regulator indicators
Содержание EVIS EXERA III
Страница 2: ......
Страница 6: ...iv Contents CLV 190 INSTRUCTION MANUAL...
Страница 20: ...14 Summary of the Equipment Functions CLV 190 INSTRUCTION MANUAL...
Страница 22: ...16 1 1 Checking the package contents list CLV 190 INSTRUCTION MANUAL Ch 1...
Страница 26: ...20 2 2 Front panel CLV 190 INSTRUCTION MANUAL Ch 2 Control panel buttons 13 9 8 7 6 5 14 12 11 10...
Страница 28: ...22 2 2 Front panel CLV 190 INSTRUCTION MANUAL Ch 2 Control panel indicators 15 18 16 26 25 23 17 22 24 20 21 19...
Страница 32: ...26 2 3 Rear and side panels CLV 190 INSTRUCTION MANUAL Ch 2...
Страница 98: ...92 6 4 Lamp usage indicator reset CLV 190 INSTRUCTION MANUAL Ch 6...
Страница 108: ...102 7 8 Transportation CLV 190 INSTRUCTION MANUAL Ch 7...