Chapter 3 Cleaning, Disinfection and Sterilization Procedures
EVIS EXERA GIF/CF/PCF TYPE 160 Series REPROCESSING MANUAL
3.6
High-level disinfection
After manual cleaning, disinfect the endoscope according to the procedures
described below.
Equipment needed
Prepare the following equipment:
All disinfection steps should be performed with the
endoscope and all equipment completely immersed. While
they are immersed, if the equipment is connected to the
endoscope or any part of them is not immersed completely,
disinfectant solution may not adequately contact all surfaces
of the equipment.
Preparation
1.
Fill a basin with disinfectant solution at the temperature and concentration
recommended by the disinfectant manufacturer. Use a basin which is at
least 40 cm by 40 cm (16” by 16”) in size and deep enough to allow the
endoscope to be completely immersed.
2.
Connect the channel plug and the injection tube to the endoscope.
3.
Attach the auxiliary water tube to the auxiliary water inlet (for endoscopes
with auxiliary water feeding only).
• Personal protective equipment
• Clean, lint-free cloths
• Large basin
• Disinfectant solution
• Channel plug
(MH-944)
• Injection tube
(MH-946)
• Auxiliary water tube
(MAJ-855 for endoscopes with
auxiliary water feeding only)