Chapter 3 Cleaning, Disinfection and Sterilization Procedures
49
EVIS EXERA GIF/CF/PCF TYPE 160 Series REPROCESSING MANUAL
Presoak for excessive bleeding and/or delayed reprocessing
Follow the steps below only in case of excessive bleeding
and/or delayed reprocessing; unnecessary immersions
should be avoided. Consecutive extended immersions may
damage the endoscope.
Preclean and perform a leakage test on the endoscope as described in Section
3.3, “Precleaning” and Section 3.4, “Leakage testing”.
1.
Fill a basin with detergent solution at the temperature and concentration
recommended by the detergent manufacturer. Use a basin which is at least
40 cm by 40 cm (16” by 16”) in size and deep enough to allow the
endoscope to be completely immersed.
2.
Carefully coil the endoscope’s insertion tube and universal cord, and
completely immerse the endoscope in the detergent solution.
Do not coil the insertion tube or universal cord with a
diameter of less than 12 cm. Equipment damage can result.
3.
Fill all the channels with the detergent solution following the procedures
described in this section.
4.
Soak the endoscope for 10 hours at the temperature recommended by the
detergent manufacturer.
5.
Remove the endoscope from the detergent solution.
6.
After soaking the endoscope, manually clean it following the procedures
described in this section, then disinfect or sterilize the endoscope following
the procedures described in Section 3.6, “High-level disinfection” or Section
3.8, “Sterilization”.